Abstract

Introduction: Percutaneous ventricular assist devices (pVADs) are increasingly used in cardiogenic shock but are associated with complications including haemolysis. The aim of this study was to investigate patient characteristics associated with haemolysis in cardiogenic shock patient population. Methods: Consecutive patients were identified using Current Procedural Terminology (CPT) codes for pVAD insertion. Patient characteristics, laboratory and imaging data, and patient outcomes were abstracted manually and using validated automated methods. Laboratory-defined haemolysis required a drop in haemoglobin ≥2 mg/dl with either lactate dehydrogenase ≥250 units/l or undetectable haptoglobin. Clinically significant haemolysis was defined as laboratory-defined haemolysis necessitating transfusion. Primary outcome was the association between haemolysis and on-device and 30-day mortality. Results: A total of 196 patients underwent pVAD insertion for cardiogenic shock during the study period and were included. Laboratory-defined haemolysis occurred in 46 patients (23.5%), of whom 12 (6.1%) had clinically significant haemolysis. Haemolysis occurred more often following emergency insertion, rather than elective insertion (84.8% versus 40.0%, p<0.001) in patients with elevated lactic acid levels (median 2.5 versus 1.6, p=0.016) and elevated heart rates (92.5 BPM versus 86.5 BPM, p=0.023). After multivariable adjustment, there was no association between laboratory-defined haemolysis and on-device (OR 0.6; 95% CI [0.1–3.4]; p=0.565) or 30-day mortality (OR 2.1; 95% CI [0.4–13.0]; p=0.391). Conclusion: Laboratory-defined haemolysis was common in patients with cardiogenic shock and pVAD, but clinically significant haemolysis was not. There was no association between haemolysis and on-device or 30-day mortality.

Disclosure:KVA has consulted for Guidepoint and Tegus and serves on a data monitoring and safety board for Fortrea. ASR has received consulting/speaking honoraria from Abiomed, Analog Devices, CareTaker Medical, Vectorious and Volumetrix and is on the Cardiac Failure Review editorial board; this did not affect peer review. All other authors have no conflicts of interest to declare.

Received:

Accepted:

Published online:

Informed Consent:

The requirement to obtain informed consent was waived by the Institutional Review Board of Vanderbilt University Medical Center.

Data Availability Statement:

The data are not publicly available due to privacy or ethical restrictions.

Ethics Approval Statement:

The study was approved by the Institutional Review Board of Vanderbilt University Medical Center.

Funding:

AES is supported by an NIH T32 award (GM007569). KVA is supported by the NIH (K23HL166960), the International Society for Heart and Lung Transplantation Enduring Hearts Transplant Longevity Award, the Red Gates Foundation, and an AHA Career Development Award (#929347).

Author contributions:

Conceptualisation: ASR; data curation: AES, HG, LD, JN, LT, EF-E; methodology: ASR; writing – original draft: AES, JLL, QSW, BDA, JMO, SB, KVA, ASR; writing – review & editing: AES, ASR.

Correspondence Details:Aniket S Rali, Division of Cardiovascular Diseases, Vanderbilt University Medical Center, 1215 21st Ave S 5th floor, Nashville, TN 37232, US. E: aniket.rali@vumc.org

Open Access:

This work is open access under the CC-BY-NC 4.0 License which allows users to copy, redistribute and make derivative works for non-commercial purposes, provided the original work is cited correctly.

Percutaneous ventricular assist devices (pVADs) are catheter-based miniaturised micro-axial flow pumps that provide circulatory support. Due to ease of placement and expanded indications, the usage of pVADs has increased recently to treat patients with various aetiologies of cardiogenic shock, including post-pericardiotomy shock.1–3 Furthermore, pVADs are used for intra-procedural circulatory support during high-risk coronary interventions and ventricular tachycardia ablation.4

Benefits of mechanical circulatory support devices must be balanced against their inherent risks. Bleeding, haemolysis, peripheral vascular complications, and stroke are among the most common complications with pVAD.5 Patients with a pVAD have higher rates of bleeding compared with other cardiac intensive care unit (ICU) patients and have nearly threefold higher rates of bleeding than those with intra-aortic balloon pump (IABP).6,7 There is also a higher incidence of haemolysis in pVAD patients compared with IABP patients and an approximately 13% risk of vascular complications, including acute limb ischaemia.8,9

Several device-specific factors, including pump speed, pump size, inlet or outlet obstruction and high shear stress along the rotor blades, lead to haemolysis.10 Many protocols currently exist to monitor for haemolysis while on pVAD support, including serial haemoglobin, lactate dehydrogenase (LDH), plasma free haemoglobin, and/or haptoglobin levels.11 However, there remains a paucity of data on patient-specific factors that increase the risk of haemolysis while on pVAD support. In this study we investigated the patient characteristics most strongly associated with both laboratory and clinically significant haemolysis in patients admitted to our cardiac ICU with pVADs.

Methods

The Vanderbilt University Medical Center (VUMC) Mechanical Circulatory Support registry is a retrospective, single-centre registry at a quaternary care hospital. The study was approved by the Institutional Review Board of Vanderbilt University Medical Center and informed consent was waived.

Study Population

We retrospectively identified all patients aged ≥18 years who underwent pVAD insertion or removal at VUMC from 1 January 2018 to 30 June 2023. All pVAD types, including Impella 2.5, CP, 5.5 and RP were included. Patients were identified using Current Procedural Terminology (CPT) codes for insertion (CPT 33990).

Data Collection and Definition of Variables

Patient demographics and comorbidity data were abstracted from the medical record system using previously established methods.12–16 Echocardiographic data, when available within 3 months prior to pVAD insertion, were abstracted from the official report. Haemodynamic and laboratory data immediately prior to pVAD insertion were extracted using the date of insertion. pVAD indication, urgency and details of the pVAD circulatory support course (including P-level, device type and insertion vessel) were determined through manual chart review.

Study Outcomes

Two definitions of haemolysis were used. Laboratory-defined haemolysis was described as a drop in haemoglobin ≥2 mg/dl and either LDH ≥250 units/l (greater than twofold the upper limit of normal) or an undetectable haptoglobin.17 Plasma free haemoglobin was obtained very infrequently and, hence, not included in this definition. Clinically significant haemolysis was defined as the presence of laboratory-defined haemolysis necessitating transfusion with at least 1 unit of packed red blood cells in the absence of an alternate source of blood loss.

Statistical Analysis

Descriptive statistics are reported for demographics, clinical characteristics and outcomes stratified by the presence or absence of laboratory haemolysis. Continuous variables are reported as median and IQR, and between-group comparisons were conducted using Mann–Whitney U-test. Categorical variables are reported as total number and percentage, with comparisons made using Pearson chi-squared test or Fisher’s exact test as appropriate. Multivariable logistic regression modelling was performed to assess the association of clinically relevant variables and the development of haemolysis. The primary outcome was the association of haemolysis with on-device and 30-day mortality, which was assessed using logistic regression models that adjusted for clinically relevant variables. Data were analysed using R version 4.2.2 (R Foundations for Statistical Computing).

Results

Baseline Cohort Characteristics

A total of 196 patients underwent pVAD insertion over the study period. The median (IQR) age of the cohort was 65.9 (55.2–75.1) years and included 60 women (30.6%) and 27 Black or Hispanic patients (13.7%; Table 1). Impella 2.5 and CP were the predominant devices implanted (89.8%) and were used in emergency implantation in 99 patients (50.5%). A total of 66 patients (33.7%) required device insertion in the setting of acute myocardial infarction (MI), while 43 (21.9%) underwent placement for heart failure-associated cardiogenic shock. The most frequent indication for implantation was bridge to recovery (40.8%), while 73 patients (37.2%) had pVAD for haemodynamic support during high-risk percutaneous coronary intervention, and 28 patients (14.3%) had pVAD for left ventricular unloading during veno-arterial extracorporeal life support.

Table 1: Baseline Characteristics of Cohort

Article image

Table 1: Baseline Characteristics of Cohort - Continued

Article image

Clinical Characteristics of Patients with Haemolysis

Of 196 patients, 46 (23.5%) met the laboratory definition for haemolysis and 12 (6.1%) met criteria for clinically significant haemolysis. Patients with laboratory-defined haemolysis were more likely to have emergency, rather than elective, pVAD insertion (84.8% versus 40.0%, p<0.001) and lower baseline left ventricular ejection fraction (LVEF; median, 18.0% versus 32.3%, p=0.002). pVAD insertion was more often in the setting of acute MI (58.7% versus 26.0%, p<0.001) and intended for recovery (67.4% versus 32.7%, p<0.001) in those with laboratory-defined haemolysis, whereas the device was more often used for haemodynamic support during high-risk procedures in those without laboratory-defined haemolysis (6.5% versus 46.7%, p<0.001). Baseline lactic acid (median, 2.5 mmol/L versus 1.6 mmol/L, p=0.016) and heart rate (92.5 BPM versus 86.5 BPM, p=0.023) were significantly higher in those with laboratory-defined haemolysis and there was a trend toward lower pulse pressure (33.5 mmHg versus 38.5 mmHg, p=0.056).

Clinical Outcomes Associated with Laboratory Haemolysis

pVAD support duration was significantly longer (67.4 hours versus 4.28 hours, p<0.001) for patients with laboratory haemolysis than for those without. Median hospital length of stay was longer in the cohort with haemolysis although it did not reach statistical significance (7.9 days versus 4.8 days, p=0.06; Table 2). Renal failure was significantly more common in those with haemolysis than in those without haemolysis (37% versus 16.7%, p=0.011).

Table 2: Outcomes Associated with Laboratory-defined Haemolysis

Article image

Predictors of Laboratory-defined Haemolysis

Clinically relevant predictors of laboratory haemolysis are listed in Table 3. In multivariable regression modelling adjusting for aetiology of cardiogenic shock, urgency of pVAD insertions, and pre-device insertion lactate, pulse pressure and LVEF, only emergency pVAD insertion remained a significant predictor of laboratory haemolysis (OR 12.3; 95% CI [1.1–137.3]; p=0.041).

Association of Laboratory Haemolysis with Mortality

The results of multivariable regression modelling examining the association between haemolysis and both on-device and 30-day mortality are listed in Table 4. After adjusting for clinically relevant variables, there was no association between laboratory-defined haemolysis and on-device mortality (OR 0.6; 95% CI [0.1–3.4]; p=0.565). There was similarly no association between laboratory-defined haemolysis and 30-day mortality (OR 2.1; 95% CI [0.4–13.0]; p=0.391).

Table 3: Predictors of Laboratory-defined Haemolysis

Article image

Table 4: Association of Laboratory-defined Haemolysis with On-device and 30-day Mortality

Article image

Clinically Significant Haemolysis

Clinically significant haemolysis was rare, occurring in 12 patients (6.1%). Baseline characteristics of those with clinically significant haemolysis are listed in Supplementary Material Table 1. Patients with clinically significant haemolysis were of similar age and gender and had similar baseline haemoglobin levels compared with the rest of the cohort. They more frequently had emergency pVAD insertion for acute MI, decompensated heart failure, or ventricular tachycardia and had longer device run times. Outcomes for these patients are summarised in Supplementary Material Table 2. Patients with clinically significant haemolysis had greater number of blood transfusions (4.0 units versus 0.0 units).

Discussion

In this single-centre retrospective study of patients receiving pVADs, we examined the impact of patient-specific factors associated with both laboratory as well as clinically significant haemolysis. The key findings of our study are, first, that rates of clinically significant haemolysis while on pVAD support were very low, although 1 in 4 patients met laboratory criteria for haemolysis; second, all of the patients with clinically significant haemolysis received either Impella 2.5 or CP device; third, laboratory-defined haemolysis was more common in patients with higher severity of presenting illness (including higher baseline lactate level, and lower LVEF and pulse pressure) requiring emergency circulatory support; and last, that haemolysis was not independently associated with higher on-device or 30-day mortality.

Haemolysis is a known complication of pVADs caused by non-laminar flow across the pump, resultant shear stress, and direct contact with mechanical device.18 Smaller pump size results in greater turbulent flow and shear stress. Higher performance levels also lead to increased likelihood of haemolysis in addition to improper device position, pump thrombosis, and longer pump run.19 In our cohort, Impella 2.5/CP pump and longer pump run times were indeed both associated with greater rates of laboratory haemolysis. The incidence of haemolysis with pVADs is reported to be between 10% and 62.5%.20,21 Rates of laboratory haemolysis in our cohort were in that range at 23.5%.

In addition to the aforementioned factors, significant variability in reported incidence of haemolysis may be driven by a lack of consistent definition of haemolysis. Traditionally, serum LDH level is used to diagnose haemolysis. The INTERMACS Registry defines major haemolysis as LDH 2.5-fold the upper limit of normal range or plasma free haemoglobin >20 mg/dl obtained 72 hours after device implantation with clinical signs of haemolysis such as jaundice, haemoglobinuria or hyperbilirubinemia.22 A major limitation of LDH as an independent marker of haemolysis is that this metabolic enzyme is found in almost all tissues in the body and can be elevated during MI, skeletal muscle damage and multi-organ failure: all common findings in patients with cardiogenic shock.23–25 In their single-centre retrospective analysis of 116 patients in cardiogenic shock, Esposito et al. found that serum LDH is significantly elevated even prior to device implantation.19 Higher acuity of cardiogenic shock at the time of device implantation was also associated with increased risk of laboratory haemolysis in our patients. In our study, patients with acute MI as the aetiology of their cardiogenic shock were indeed more likely to meet criteria for laboratory haemolysis. Although plasma free haemoglobin has been shown to be a more reliable marker of haemolysis in patients with pVAD, it was not included in our definition of haemolysis because this was not routinely checked at our institution during the study period. Instead, low haptoglobin was used as an additional marker of haemolysis given its propensity to bind free haemoglobin.19,26

Our study found that emergency pVAD insertion was an independent risk factor for laboratory-defined haemolysis. It is plausible that this may be driven by a higher likelihood of improper positioning with emergency placement, which has been shown to be associated with increased incidence of haemolysis.27 Alternatively, emergency placement is more likely to be performed in a sicker patient cohort requiring higher performance levels.18,19,28 pVAD support as a bridge to recovery was also associated with higher rates of haemolysis in our cohort and is likely to have been due to longer exposure to the device.29

Renal injury is a common consequence of intravascular haemolysis. By-products of red blood cell breakdown, such as haem, are directly cytotoxic to the proximal tubule and lead to renal vasoconstriction, which can precipitate an acute kidney injury.30 High rates of renal failure have been confirmed in patients supported with pVAD.21 This finding is often confounded by the propensity for patients in cardiogenic shock to develop acute renal failure regardless of mechanical support. In our cohort, renal failure was more common in patients with laboratory-defined haemolysis (37% versus 16%, p=0.011), but the same patients had more severe cardiogenic shock, as indicated by the higher serum lactate levels and lower pulse pressure and LVEF. It is important to note that although laboratory haemolysis may contribute to this comorbidity for patients on pVAD support, it was not associated with on-device or 30-day mortality. It is likely that the risk of such comorbidities is offset by the mortality benefit from restoration of adequate circulatory support, as was seen in the recently published DanGer Shock trial.31

There are a few notable limitations of our study. First, this is a retrospective analysis and is subject to all limitations inherent in such a design. Second, this is a single-centre cohort, which limits the generalisability of our findings. Lastly, we attempted to adjust for all clinically relevant confounders, but we appreciate that not all potential confounders may have been captured in our database. For example, at our institution it is standard practice for all patients with a pVAD to be on a heparin drip titrated to an unfractionated level of 0.2–0.4 IU/ml, but we are unable to quantify time in the therapeutic range.

Conclusion

Laboratory-defined haemolysis was common in patients with pVADs, but clinically significant haemolysis was relatively uncommon. There are many potential confounders when using a laboratory-defined diagnosis of haemolysis in the setting of pVAD. Haemolysis on pVAD was not associated with on-device or 30-day mortality in our cohort.

Click here to view Supplementary Material.

Clinical Perspective

  • Plasma free haemoglobin may be a more reliable marker of haemolysis in Impella patients, given that lactate dehydrogenase may be elevated from non-red blood cell turnover in clinically ill patients.
  • The incidence of laboratory haemolysis increases as the severity of acute illness increases in critically ill patients.

References

  1. Amin AP, Spertus JA, Curtis JP, et al. The evolving landscape of Impella use in the United States among patients undergoing percutaneous coronary intervention with mechanical circulatory support. Circulation 2020;141:273–84. 
    Crossref | PubMed
  2. Kim Y, Shapero K, Ahn SS, et al. Outcomes of mechanical circulatory support for acute myocardial infarction complicated by cardiogenic shock. Catheter Cardiovasc Interv 2022;99:658–63. 
    Crossref | PubMed
  3. Fried J, Farr M, Kanwar M, et al. Clinical outcomes among cardiogenic shock patients supported with high-capacity Impella axial flow pumps: a report from the Cardiogenic Shock Working Group. J Heart Lung Transplant 2024;43:1478–88. 
    Crossref | PubMed
  4. Agrawal A, Shoemaker J, Bommareddi S, Dumitru I. Impella RP flex: rescue for the failing right ventricle after heart transplantation: case report. ASAIO J 2025;71:e33–7. 
    Crossref | PubMed
  5. Saito Y, Tateishi K, Toda K, et al. Complications and outcomes of Impella treatment in cardiogenic shock patients with and without acute myocardial infarction. J Am Heart Assoc 2023;12:e030819. 
    Crossref | PubMed
  6. Scherer C, Lüsebrink E, Kupka D, et al. Long-term clinical outcome of cardiogenic shock patients undergoing Impella CP treatment vs. standard of care. J Clin Med 2020;9:3803. 
    Crossref | PubMed
  7. Schrage B, Ibrahim K, Loehn T, et al. Impella support for acute myocardial infarction complicated by cardiogenic shock. Circulation 2019;139:1249–58. 
    Crossref | PubMed
  8. Pieri M, Sorrentino T, Oppizzi M, et al. The role of different mechanical circulatory support devices and their timing of implantation on myocardial damage and mid-term recovery in acute myocardial infarction related cardiogenic shock. J Interv Cardiol 2018;31:717–24. 
    Crossref | PubMed
  9. Patel N, Sharma A, Dalia T, et al. Vascular complications associated with percutaneous left ventricular assist device placement: a 10-year US perspective. Catheter Cardiovasc Interv 2020;95:309–16. 
    Crossref | PubMed
  10. Fishkin T, Isath A, Naami E, et al. Impella devices: a comprehensive review of their development, use, and impact on cardiogenic shock and high-risk percutaneous coronary intervention. Expert Rev Cardiovasc Ther 2023;21:613–20. 
    Crossref | PubMed
  11. Van Edom CJ, Gramegna M, Baldetti L, et al. Management of bleeding and hemolysis during percutaneous microaxial flow pump support: a practical approach. JACC Cardiovasc Interv 2023;16:1707–20. 
    Crossref | PubMed
  12. Siddiqi HK, Trahanas J, Xu M, et al. Outcomes of heart transplant donation after circulatory death. J Am Coll Cardiol 2023;82:1512–20. 
    Crossref | PubMed
  13. Beckman JA, Duncan MS, Damrauer SM, et al. Microvascular disease, peripheral artery disease, and amputation. Circulation 2019;140:449–58. 
    Crossref | PubMed
  14. Davogustto GE, Clark DE, Hardison E, et al. Characteristics associated with multisystem inflammatory syndrome among adults with SARS-CoV-2 infection. JAMA Netw Open 2021;4:e2110323. 
    Crossref | PubMed
  15. Bielinski SJ, Pathak J, Carrell DS, et al. A robust e-epidemiology tool in phenotyping heart failure with differentiation for preserved and reduced ejection fraction: the electronic medical records and genomics (eMERGE) network. J Cardiovasc Transl Res 2015;8:475–83. 
    Crossref | PubMed
  16. Shuey MM, Stead WW, Aka I, et al. Next-generation phenotyping: introducing phecodeX for enhanced discovery research in medical phenomics. Bioinformatics 2023;39:btad655. 
    Crossref | PubMed
  17. Ravichandran AK, Parker J, Novak E, et al. Hemolysis in left ventricular assist device: a retrospective analysis of outcomes. J Heart Lung Transplant 2014;33:44–50. 
    Crossref | PubMed
  18. Yasuda T, Shimokasa K, Funakubo A, et al. An investigation of blood flow behavior and hemolysis in artificial organs. ASAIO J 2000;46:527–31. 
    Crossref | PubMed
  19. Esposito ML, Morine KJ, Annamalai SK, et al. Increased plasma-free hemoglobin levels identify hemolysis in patients with cardiogenic shock and a trans valvular micro-axial flow pump. Artif Organs 2019;43:125–31. 
    Crossref | PubMed
  20. Zaiser AS, Fahrni G, Hollinger A, et al. Adverse events of percutaneous microaxial left ventricular assist devices: a retrospective, single-centre cohort study. J Clin Med 2021;10:3710. 
    Crossref | PubMed
  21. Badiye AP, Hernandez GA, Novoa I, Chaparro SV. Incidence of hemolysis in patients with cardiogenic shock treated with Impella percutaneous left ventricular assist device. ASAIO J 2016;62:11–14. 
    Crossref | PubMed
  22. Papanastasiou CA, Kyriakoulis KG, Theochari CA, et al. Comprehensive review of hemolysis in ventricular assist devices. World J Cardiol 2020;12:334–41. 
    Crossref | PubMed
  23. Zhang H, Kang K, Chen S, et al. High serum lactate dehydrogenase as a predictor of cardiac insufficiency at follow-up in elderly patients with acute myocardial infarction. Arch Gerontol Geriatr 2024;117:105253. 
    Crossref | PubMed
  24. Reis GJ, Kaufman HW, Horowitz GL, Pasternak RC. Usefulness of lactate dehydrogenase and lactate dehydrogenase isoenzymes for diagnosis of acute myocardial infarction. Am J Cardiol 1988;61:754–8. 
    Crossref | PubMed
  25. Akasha R, Amanullah M, Ali SP, et al. Assessment of acute myocardial infarction by the use of special biochemical markers. Ulutas Med J 2015;1:68−73. 
    Crossref
  26. Körmöczi GF, Säemann MD, Buchta C, et al. Influence of clinical factors on the haemolysis marker haptoglobin. Eur J Clin Invest 2006;36:202–9. 
    Crossref | PubMed
  27. Roberts N, Chandrasekaran U, Das S, et al. Hemolysis associated with Impella heart pump positioning: in vitro hemolysis testing and computational fluid dynamics modeling. Int J Artif Organs 2020;43:391398820909843. 
    Crossref | PubMed
  28. Cannata A, Cantoni S, Sciortino A, et al. Mechanical hemolysis complicating transcatheter interventions for valvular heart disease: JACC state-of-the-art review. J Am Coll Cardiol 2021;77:2323–34. 
    Crossref | PubMed
  29. Panoulas V, Monteagudo-Vela M. Predictors of short-term survival in cardiogenic shock patients requiring left ventricular support using the Impella CP or 5.0. CJC Open 2021;3:1002–9. 
    Crossref | PubMed
  30. Van Avondt K, Nur E, Zeerleder S. Mechanisms of haemolysis-induced kidney injury. Nat Rev Nephrol 2019;15:671–92. 
    Crossref | PubMed
  31. Møller JE, Engstrøm T, Jensen LO, et al. Microaxial flow pump or standard care in infarct-related cardiogenic shock. N Engl J Med 2024;390:1382–93. 
    Crossref | PubMed